PSUR Under EU MDR: Preparing the Report and Staying Compliant

PSUR Under EU MDR: Preparing the Report and Staying Compliant

Once a medical device is on the market, the manufacturer’s job isn’t finished. The Periodic Safety Update Report (PSUR) is how manufacturers show EU MDR PSUR, on a regular basis, that their device is still safe and performing as intended. Under Article 86 of Regulation (EU) 2017/745, manufacturers of Class IIa, IIb, and III devices must prepare one, based on their post-market surveillance (PMS) data.

1. Know what the PSUR is meant to do

The PSUR brings together everything your post-market surveillance has uncovered and records any preventive or corrective actions you’ve taken. Its central purpose is to answer one question: does the device still offer an acceptable balance of benefits and risks across its expected lifetime?

2. Work out how often to update it

How frequently you need to update the PSUR depends on the device class:

  • Class IIa: at least once every two years

  • Class IIb: at least once a year

  • Class III: at least once a year

Class I devices follow a different route. Instead of a PSUR, they need a Post-Market Surveillance Report (PMSR) under Article 85.

3. Collect and analyse your post-market data

Start by gathering information from all the places real-world feedback shows up:

  • Customer complaints and user feedback

  • Serious incidents and vigilance reports

  • Field safety corrective actions

  • Post-market clinical follow-up (PMCF) activities

  • Sales volumes and estimated user numbers

  • Trends and any newly emerging safety concerns

Collecting the data is only half the task. The real value comes from analysing it: spotting patterns, noticing problems that are becoming more common, and flagging anything your original risk assessment didn’t foresee.

4. Put the report together

A complete PSUR should cover:

  • The conclusions of your benefit-risk determination

  • The main findings from your PMCF

  • Sales volume and the estimated size of the user population

  • The rationale for, and a description of, any preventive and corrective actions

Review and Maintain Compliance

Preparing an accurate EU MDR PSUR helps manufacturers monitor device safety, evaluate clinical performance, and maintain regulatory compliance. Regular data analysis and timely report updates support effective post-market surveillance throughout the medical device lifecycle.

Book a Demo