How to Prepare a GSPR Checklist for Your Medical Device

How to Prepare a GSPR Checklist for Your Medical Device

Before a medical device can be sold in Europe, the manufacturer has to show it meets the General Safety and Performance Requirements (GSPR) set out in Annex I of Regulation (EU) 2017/745 (the EU MDR). The GSPR checklist is the document that ties it all together: it shows, requirement by requirement, how your device measures up and where the proof lives.

Step 1: Get to Know Annex I

Start by reading Annex I properly. It covers a lot of ground: general safety, risk management, design and manufacturing, chemical and biological properties, infection and microbial contamination, and the information you supply with the device. You won’t need every requirement, but you do need to know what’s there before deciding what applies.

Step 2: Decide What Applies to Your Device

Look at your device’s intended purpose, classification, design, and characteristics, then go through Annex I and mark each requirement as applicable or not applicable. If you mark something as not applicable, write down why. A clear justification saves you from awkward questions later.

Step 3: Gather Your Evidence

For every applicable requirement, you’ll need objective proof that you meet it. Depending on your device, that might include: Risk management reports Clinical evaluation reports Biocompatibility assessments Electrical safety and performance test results Software verification and validation records Sterilisation validation, if the device is supplied sterile Labelling and instructions for use Step 4: Map Evidence to Each Requirement Now build the checklist itself. A good one includes, for each requirement: The GSPR reference number A description of the requirement Whether it applies Your conformity status The supporting evidence The specific document references Laid out this way, gaps are easy to spot and every claim can be traced back to its source.

Step 5: Keep It Current

A GSPR checklist is a living document. Revisit it whenever your design, intended purpose, applicable standards, or the regulations change, and make sure it stays consistent with your technical documentation and risk management records. An outdated checklist can be worse than none at all, because it contradicts the rest of your file.

Review and Update the Checklist

A well-built GSPR checklist medical devices makes technical documentation reviews and conformity assessments much smoother. With accurate evidence and clear traceability, you can show a notified body exactly how your device meets the requirements, without scrambling to find the paperwork.